SDS-PAGE异常电泳现象及分析

萌到你眼炸
966次浏览
2020年07月30日 15:24
最佳经验
本文由作者推荐

预备党员考察意见-元旦祝福




SDS-PAGE异常电泳现象及分析
(SDS-PAGE “Hall of Shame”)

本资料来源于生物秀(网络连接地址:
http:


You may very well have prepared a nearly perfect gel, and would have a difficult time
improving upon the product. If that is so, then by all means gloat about it! If you didn't do
such a hot job don't despair. All good scientists learn from their mistakes, in fact, without
making mistakes most of us wouldn't learn much at all!
The
have run in past labs, with an example or two from a research lab. They represent many of
the ways one can mess up a gel (but not all of them - we're still finding new ways!). See
what features of your own gel(s) were unsatisfactory - or at least less than perfect - and
use the illustrations to figure out what you might do to improve your technique.
To critique your own work identify your symptoms and use the gallery to select
appropriate example(s). Each example is linked to a full sized image with suggessted
explanations for the symptoms. From this
what can be done to correct the problem(s). If you wish you may browse by descriptions of
symptoms, listed below the gallery.
Symptoms 1—Smears
Smeared gels – example 1
Smearing can have a variety of causes, but most commonly it is due to an unevenly
poured acrylamide mixture or due to gross overloading of protein.



In this example the gel was not properly poured, so that the lower half had begun to
polymerize before the upper part was poured. The first gel mix began to polymerize too
quickly. Rather than prepare a new gel cassette, the students simply stopped pouring,
prepared a new mix, and poured it on top of the old. Obviously, the bond wasn't
particularly good.
Smeared gels – example 2
In this example a lot of protein was loaded in each of the wells, all the way across. Most of
the lanes can still be interpreted, but smearing is particularly evident in lanes 3 and 4.
Those lanes contained primarily one protein (hemoglobin subunit), as did lanes 9 and 10.
However, the group that loaded 3 and 4 underestimated the protein concentrations in their
red cell lysate and red cell cytoplasm fractions. Those fractions contained so much protein
that samples had to be diluted prior to preparing a protein assay tube. The students forgot
to take the dilution factor into consideration when determining protein concentration.
The capacity of a mini-gel for a mixed protein sample is 20 to 40 microgramswell,
depending on the resolution needed and number individual polypeptides in the mix.
However, if a pure protein is loaded, one single band will contain all 20-40 micrograms,
and the result is a mess.




Smeared gels – example 3
Here is another example of gross overloading, and it appears these students made the
same mistake as in example 2. It looks as though there was some inconsistency in the gel
also.

Smeared gels – example 4
Smearing can be a normal consequence of running membrane-associated proteins with a
high lipid content in a discontinuous gel. In discontinuous PAGE, sample proteins are
super- concentrated by a combination of two factors. First, the rate of migration is suddenly
slowed as the sample moves from a non- restrictive stacking gel to a separating gel that
restricts movement. Second (and more important), a pH change affects the


electrophoretic mobility of proteins in the sample, causing the sample to check up abruptly.
A sample of a half cm or so in height becomes compressed into a layer of bands a few
micrometers thick, and the local concentration of protein goes up considerably.
Membrane-associated proteins tend to precipate at lower concentrations, thus lanes
containing such proteins often have a dark band of precipitated proteins at the very top.
As electrophoresis proceeds the precipates re- dissolve and enter the gel continuously,
thus forming a continuous dark background of unresolved polypeptides. Many of the
membrane samples shown on these gels will show that characteristic.

The resolved bands in the membrane protein lanes (4 and 7) appear to consist of similar
amounts of protein, however in lane 4 the capacity of the gel for at least some of the
membrane proteins was exceeded, resulting in a dark smear.
Symptoms 2—Streaks
Streaks – example 1
This one isn't so bad. The lane with the streaks (left of lane 5) is quite usable, in fact. The
streaks came about because a portion of the more concentrated proteins in the


membrane fraction precipated during the stacking process, then returned to solution after
a fairly short time. Because of slight imperfections in the separating gel surface (it may
have been allowed to dry a little before adding the stacking gel), the protein concentration
was not uniform across the lane, and streaks rather than a uniform darker background are
the result.

Streaks – example 2



The fourth lane is streaked due to the effect described above. The electrical field was
affected by the non- uniform concentration in the fourth lane (the dark one with streaks),,
causing a narrowing of the bands. Note that the dye front was lost in this gel - it was
allowed to run too long.
Streaks – example 3

This one kind of stinks. Evidently, this group did not use an overlay after pouring the first
gel layer. The result was a separating gel with the very rough top that you see here. As
samples were stacked they were disproportionately concentrated in the 'valleys,' where
the proteins precipitated. The precipitates re-dissolved gradually during the run, forming


streaks instead of resolving into discrete bands.

Symptoms 3—Bands too light
Light bands – example 1
It looks like this problem was caused by badly formed wells andor careless loading of
sample onto the gel. The background is dark and uneven, suggesting that protein was in
the running buffer itself. Most likely the apparatus was jarred and much of the samples
was slopped out of the wells. The spilled sample would not resolve into bands,since it


continuously entered the gel from the upper buffer compartment.

Note the presence of a horizontal line, continuous along several lanes (arrow). That would
suggest that the wells were too shallow, and some material spilled into adjacent wells.
This symptom also shows up in perfectly good gels when the electrodes are hooked up
backwards for a few minutes. Protein migrates out of the wells instead of into the stacking
gel. Catching the problem within a few minutes salvages some of the information, but the
pH effect that causes efficient stacking is compromised, sample is lost, and protein
contaminates the running buffer.
Light bands – example 2
Judging from the lack of other symptoms, it appears that the gel simply didn't stain well.
Coomassie blue dye staining solution can become contaminated with SDS if it is recycled.
The dye becomes less effective and proteins don't show up as well as with fresh dye. As
long as the proteins were precipitated in the gel by the acidified alcohol in the stain


solution, the problem could be corrected simply by re-staining with fresh dye.

Light bands – example 3
The 'X' marks the lane next to lane 5 where the amount of protein was overestimated in
the original sample, the sample was prepared to the wrong concentration for
electrophoresis, or too little was loaded - perhaps sample wasn't properly drawn up into
the syringe. The absence of any major bands, compared to the other samples, suggests


that the well was simply underloaded.

Light bands – example 4
This is one of the more disappointing symptoms you might encounter. Let's say you were
extremely meticulous,did everything perfectly, and ran your gel so that the dye front was
perfectly even and right at the bottom. You rinsed the gel in deionized water, then left it on
the shaker. Being late in the day, you forgot to replace the water with the stain. It was
stained the next day, but here's the result:



Notice that the smaller proteins diffused rather quickly out of the gel, while the larger ones
at the top diffused much more slowly, and took the stain. The acidified alcohol in the stain
solution and in the destain solutions is essential, as it precipates proteins, preventing them
from diffusing out of the acrylamide matrix.
Symptoms 4—Missing Dye Front
Partial dye front – example 1
After doing so much work getting the gel set up, it is a shame to be late taking it down.
Relative mobilities cannot be determined with accuracy if the dye front is partially off of the
gel.



In this case the gel was internally calibrated - that is, molecular weight standards were
included on the gel. An arbitrary reference point can be used for setting up the standard
curve, since the lanes were quite even.
No dye front – example 2
This one was allowed to run until the dye front completely ran off. Since the lanes are
somewhat distorted,relative mobility measurement will be compromised. With an intact
dye front, lanes could be compared even if the gel was distorted or of different height on
one side or the other. Relative migration is the same for a gel of uniform composition, no


matter what the actual height of the resolving gel.

No dye front – example 3
The curvature of this gel suggests that the spacers on the sides were a bit loose, causing
a field effect near the edges (a 'frowning' gel). It will be difficult to get accurate relative
mobility measurements because the dye front was allowed to run off. With an intact dye
front as a reference point, lanes could be compared with one another despite the 'frown.'


Symptoms 5—Bands only at top of gel
Stopped too soon – example 1
It is obvious where the dye front was on this gel. There was room to continue
electrophoresis, which would have spread out the bands even more and improved
resolution. Note that the pattern is consistent with the other gels that
Relative mobility, for gels of identical composition, is the same for identical proteins. The
resolution is improved with the distance the proteins are permitted to migrate, until
resolution becomes limited by diffusion.

Stopped too soon – example 2
Each of these gels – one of about 7%T and one about 14%T – was looking good.
Someone became impatient and stopped the run early. They are usable, but the
separation would have been much better if the dye front had been allowed to reach the
bottom of the gel. This will happen when you run late, the instructor gets hungry, and


you aren't there to keep him from taking down your gel.


Symptoms 6—Miscellaneous
Failed Gels - Crooked, Broken, Smeared, Weird
Crooked bands
Examples of this problem are scattered throughout the 'hall of shame.' The surface of the
resolving gel was uneven, so that when the samples were stacked, the bands started out
with distorted shapes.



Crooked bands also result from an uneven electric field, but usually that type of distortion
is symmetrical.
Bad wells
If samples run over into adjacent wells, you will see what looks like a continuous protein
band running across the wells (arrow). In fact, that's exactly what it is. This is usually what
happens when the wells are too short, or perhaps sample is accidentally spilled outside
the wells.
In addition, we may have used old sample buffer. An advantage of using Cleland's
reagent (dithithreitol, or DTT) to reduce disulfide bonds is that it doesn't stink as badly as
2-mercaptoethanol. However, the latter compound is much more stable. Disulfide bonds
link individual polypeptides together and maintain some proteins in a folded state under


otherwise denaturing conditions. This causes them to run unpredictably.

Dense Gel
...or should it be dense investigators? The gel mix was clearly incorrect. The acrylamide
concentration was so high that only the smallest proteins entered the gel to any extent.
The bands were distorted due to overloading and the fact that beyond a reasonable


concentration the most concentrated gels tend not to polymerize evenly.

Broken Gel
This can be really frustrating. You did a very nice job, then the gel broke during handling.
Keep in mind that the gel will still be usable as long as you save the pieces.


Field Effects
To some extent this is normal, although for some reason the field effect was exaggerated
here. The spacers interrupt the electric field at the edges of the gel, so that the 'pull' (or
'push' if you prefer) is biased to one side the closer the sample is to the edge of the gel.
The result is the 'frowning'effect you see here.

The individual bands sometimes smile, but if the gels have any expression at all, they
always frown. I guess you would too if we zapped you with electricity and drowned you in
blue dye
Symptoms 7—Failed Gels with Multiple Symptoms

Multiple symptoms – example 1

Many times the gels you run will have a variety of symptoms. Since the object of a critique
is to improve your work, it is important to determine what caused the undesirable effects.
Then you can take steps to improve the quality of your gels.



This one has no dye front - electrophoresis was terminated too late. Bands are crooked -
the top of the resolving gel was uneven, most likely due to failure to use a properly overlay.
On the right side, samples were overloaded.
Multiple symptoms – example 2
This gel shows some field effects (curvature - lanes are not straight). There is some


overloading, and some bands are crooked due to an uneven resolving gel surface.

Multiple symptoms – example 3
Uneven loading (different amounts in adjacent wells) has caused some compression in
the standards lane. There are tears in the gel and discoloration throughout. This indicates
that it was not properly handled - handling a gel with bare hands leaves protein on the


surface that stains with Coomassie. Sometimes you get very nice fingerprints.

Symptoms 8—Bizarre Results Explained
Bizarre – example 1
The large container of glycine looked the same as the large container of choline chloride.
Choline is a lot more hydroscopic than glycine, which should have been a clue that the
wrong component was used in the electrode buffer. Choline is a basic compound and
certainly does not substitute for glycine. We caught the error and replaced the buffer after


electrophoresis had proceeded for some time, but it was too late for this one.

Bizarre – example 2
This one was kindly contributed by T. Sedlacek, Czech Republic. It was run with high


current and overheated

Bizarre – example 3
Here are two gels that were contributed by E. Morales Rayas. A likely explanation is that
each time there was a delay between loading the samples and actually running the gel.
The middle lanes of the first gel show an alternating pattern in which every other lane is
either wide or compressed. Bands on the second gel spread out toward the edges.
When samples sit in a well the proteins begin to diffuse into the stacking gel, both
vertically and laterally. Smaller proteins diffuse more rapidly than do larger ones. If
proteins in a sample diffuse laterally they may alter the electric field affecting adjacent
lanes, especially if the samples in adjacent wells contain predominantly higher mass
polypeptides. The problem isn't so bad when all of the samples are of similar composition


(second gel) and the lanes are loaded in sequence. However distortion still occurs.


Bizarre – example 4
Evidently it is not a good idea to store tris- glycine electrode buffer in a gallon plastic bottle
on a shelf in the didn't see any mold in the buffer, but by golly it was there, all right.


Notice that in each example the gel below the dye front was cleared following destaining.
The persistent blue background is from mold proteins that continued to penetrate the gel
throughout the procedure.



公安部出入境管理局-吉林工商学院网站


硅酸盐研究所-司考真题


陕西高考人数-北京国际商务学院


南京工业大学浦江学院-写给领导的感谢信


李嘉诚励志名言-英文邀请信


爱的奉献作文-教师节的祝福


英国时间查询-邀请函内容


母亲节英语-庆祝教师节手抄报